Product Overview
OP3303 handhold optical variable attenuator is used for continuously variable optical signal attenuation. As the attenuator is used in the laser system for the on-line testing, therefore, OP3303 can be used in the digital system of communication devices (such as: PHD, SDH) and also in the system of adopting analog modulation (CATV)
Product Features
(1) Stepwise attenuating by circum gyrated dial: attenuating step 0.05dB
(2) Provide with the function of displaying dB and dBm attenuating value
(3) 10 minutes Auto-off function can be activate and deactivate with keypad operation.
(4) After off the instruments, the system will have the memorizing of the attenuating value and the attenuating step, in order to restore the system back to the previous shut down State when open the instruments next time.
(5) Portable, rugged, lightweight; Easy to use.
Application
(1) Telecom Maintenance
(2) CATV Maintenance
(3) Comprehensive cable construction system
(4) Optical instruments research and development
(5) Optical communication education and lab testing
(6) Other optical project
Specifications
Type |
OP3303 |
Attenuating wavelength Range |
1260~1650nm |
Fiber Model |
9/125um SM |
Optical Connector |
FC/PC |
Calibrated wavelengths |
1310/1490/1550/1625 |
Measurement Range |
2.5~60dB |
Resolution |
0.05dB |
Minimize Insertion Loss |
≌2.5dB |
Linearity |
∮0.5dB |
Repeating |
∮0.2dB |
Attenuating Accuracy |
∮0.8dB |
Back Reflection |
≈35dBtypical value40dB |
Max input |
<20dBm |
Displaying type |
lattice 128*64 black and white, white back ground light |
Rechargeable batteries |
7.4V |
Power supply adaptor |
7~8.5V |
Operation temperature |
0~40 |
Storage temperature |
-10~60 |
humidity |
0~85%non- condensation |
Weight |
450g |
Standard Packages
MODEL |
INCLUDES |
OP3303 |
OP3303 Main Body, Protective Rubber Boot, Rechargeable battery, Power Supply Adaptor, Instruction Manual and Cotton Swabs and Rigid hard carrying case. | |